Bias-variance tradeoff in anticorrelated noise reduction for spectral CT
KTH, Skolan för teknikvetenskap (SCI), Fysik, Medicinsk bildfysik.ORCID-id: 0000-0002-5092-8822
KTH, Skolan för teknikvetenskap (SCI), Fysik, Medicinsk bildfysik.
2017 (engelsk)Inngår i: Medical physics (Lancaster), ISSN 0094-2405, Vol. 44, nr 9, s. E242-E254Artikkel i tidsskrift (Fagfellevurdert) Published
Abstract [en]

Purpose: In spectral CT, basis material decomposition is commonly used to generate a set of basis images showing the material composition at each point in the field of view. The noise in these images typically contains anticorrelations between the different basis images, which leads to increased noise in each basis image. These anticorrelations can be removed by changing the basis functions used in the material decomposition, but the resulting basis images can then no longer be used for quantitative measurements. Recent studies have demonstrated that reconstruction methods which take the anticorrelations into account give reduced noise in the reconstructed image. The purpose of this work is to analyze an analytically solvable denoising model problem and investigate its effect on the noise level and bias in the image as a function of spatial frequency. Method: A denoising problem with a quadratic regularization term is studied as a mathematically tractable model for such a reconstruction method. An analytic formula for the resulting image in the spatial frequency domain is presented, and this formula is applied to a simple mathematical phantom consisting of an iodinated contrast agent insert embedded in soft tissue. We study the effect of the denoising on the image in terms of its transfer function and the visual appearance, the noise power spectrum and the Fourier component correlation coefficient of the resulting image, and compare the result to a denoising problem which does not model the anticorrelations in the image. Results: Including the anticorrelations in the noise model of the denoising method gives 3-40% lower noise standard deviation in the soft-tissue image while leaving the iodine standard deviation nearly unchanged (0-1% difference). It also gives a sharper edge-spread function. The studied denoising method preserves the noise level and the anticorrelated structure at low spatial frequencies but suppresses the noise and removes the anticorrelations at higher spatial frequencies. Cross-talk between images gives rise to artifacts at high spatial frequencies. Conclusions: Modeling anticorrelations in a denoising problem can decrease the noise level in the basis images by removing anticorrelations at high spatial frequencies while leaving low spatial frequencies unchanged. In this way, basis image cross-talk does not lead to low spatial frequency bias but it may cause artifacts at edges in the image. This theoretical insight will be useful for researchers analyzing and designing reconstruction algorithms for spectral CT.

sted, utgiver, år, opplag, sider
WILEY , 2017. Vol. 44, nr 9, s. E242-E254
Emneord [en]
anticorrelated noise, basis material decomposition, computed tomography, denoising, spectral CT
HSV kategori
Identifikatorer
URN: urn:nbn:se:kth:diva-217065DOI: 10.1002/mp.12322ISI: 000411804500014PubMedID: 28901607Scopus ID: 2-s2.0-85029324082OAI: oai:DiVA.org:kth-217065DiVA, id: diva2:1159402
Konferanse
4th International Conference on Image Formation in X-Ray Computed Tomography, JUL 18-22, 2016, Bamberg, GERMANY
Merknad

QC 20171122

Tilgjengelig fra: 2017-11-22 Laget: 2017-11-22 Sist oppdatert: 2017-11-22bibliografisk kontrollert

Open Access i DiVA

Fulltekst mangler i DiVA

Andre lenker

Forlagets fulltekstPubMedScopushttp://www.dsct.com/index.php/4th-international-conference-on-image-formation-in-x-ray-computed-tomography/

Personposter BETA

Persson, Mats

Søk i DiVA

Av forfatter/redaktør
Persson, MatsGrönberg, Fredrik
Av organisasjonen
I samme tidsskrift
Medical physics (Lancaster)

Søk utenfor DiVA

GoogleGoogle Scholar

doi
pubmed
urn-nbn

Altmetric

doi
pubmed
urn-nbn
Totalt: 17 treff
RefereraExporteraLink to record
Permanent link

Direct link
Referera
Referensformat
  • apa
  • ieee
  • modern-language-association-8th-edition
  • vancouver
  • Annet format
Fler format
Språk
  • de-DE
  • en-GB
  • en-US
  • fi-FI
  • nn-NO
  • nn-NB
  • sv-SE
  • Annet språk
Fler språk
Utmatningsformat
  • html
  • text
  • asciidoc
  • rtf