Inter-laboratory validation of bioaccessibility testing for metals
Vise andre og tillknytning
2014 (engelsk)Inngår i: Regulatory toxicology and pharmacology, ISSN 0273-2300, E-ISSN 1096-0295, Vol. 70, nr 1, s. 170-181Artikkel i tidsskrift (Fagfellevurdert) Published
Abstract [en]

Bioelution assays are fast, simple alternatives to in vivo testing. In this study, the intra- and inter-laboratory variability in bioaccessibility data generated by bioelution tests were evaluated in synthetic fluids relevant to oral, inhalation, and dermal exposure. Using one defined protocol, five laboratories measured metal release from cobalt oxide, cobalt powder, copper concentrate, Inconel alloy, leaded brass alloy, and nickel sulfate hexahydrate. Standard deviations of repeatability (S-r) and reproducibility (S-R) were used to evaluate the intra- and inter-laboratory variability, respectively. Examination of the s(R):s(r) ratios demonstrated that, while gastric and lysosomal fluids had reasonably good reproducibility, other fluids did not show as good concordance between laboratories. Relative standard deviation (RSD) analysis showed more favorable reproducibility outcomes for some data sets; overall results varied more between- than within-laboratories. RSD analysis of s(r) showed good within-laboratory variability for all conditions except some metals in interstitial fluid. In general, these findings indicate that absolute bioaccessibility results in some biological fluids may vary between different laboratories. However, for most applications, measures of relative bioaccessibility are needed, diminishing the requirement for high inter-laboratory reproducibility in absolute metal releases. The inter-laboratory exercise suggests that the degrees of freedom within the protocol need to be addressed.

sted, utgiver, år, opplag, sider
2014. Vol. 70, nr 1, s. 170-181
Emneord [en]
Metals, Alloys, UVCBs, Classification, Bioelution, Bioaccessibility, Read-across, Inter-laboratory validation
HSV kategori
Identifikatorer
URN: urn:nbn:se:kth:diva-154753DOI: 10.1016/j.yrtph.2014.06.021ISI: 000342482200018Scopus ID: 2-s2.0-84904408802OAI: oai:DiVA.org:kth-154753DiVA, id: diva2:763118
Merknad

QC 20141113

Tilgjengelig fra: 2014-11-13 Laget: 2014-10-27 Sist oppdatert: 2018-01-11bibliografisk kontrollert

Open Access i DiVA

Fulltekst mangler i DiVA

Andre lenker

Forlagets fulltekstScopus

Personposter BETA

Herting, GunillaWallinder, Inger Odnevall

Søk i DiVA

Av forfatter/redaktør
Herting, GunillaWallinder, Inger Odnevall
Av organisasjonen
I samme tidsskrift
Regulatory toxicology and pharmacology

Søk utenfor DiVA

GoogleGoogle Scholar

doi
urn-nbn

Altmetric

doi
urn-nbn
Totalt: 98 treff
RefereraExporteraLink to record
Permanent link

Direct link
Referera
Referensformat
  • apa
  • harvard1
  • ieee
  • modern-language-association-8th-edition
  • vancouver
  • Annet format
Fler format
Språk
  • de-DE
  • en-GB
  • en-US
  • fi-FI
  • nn-NO
  • nn-NB
  • sv-SE
  • Annet språk
Fler språk
Utmatningsformat
  • html
  • text
  • asciidoc
  • rtf