Evaluation of tissue Doppler-based velocity and deformation imaging: a phantom study of ultrasound systems.
KTH, Skolan för teknik och hälsa (STH), Medicinsk teknik, Medicinsk bildteknik.ORCID-id: 0000-0001-9419-910X
KTH, Skolan för teknik och hälsa (STH), Medicinsk teknik.
KTH, Skolan för teknik och hälsa (STH), Medicinsk teknik.ORCID-id: 0000-0002-1240-2555
2011 (Engelska)Ingår i: European Journal of Echocardiography, ISSN 1525-2167, E-ISSN 1532-2114, Vol. 12, nr 6, s. 467-476Artikel i tidskrift (Refereegranskat) Published
Abstract [en]

AIMS: The objective of this study was to test the accuracy and diagnostic interchangeability of tissue Doppler-based displacement, velocity, strain, and strain rate measurements in commercially used ultrasound (US) systems. METHODS AND RESULTS: Using an in-house made phantom, four different US scanner models were evaluated. Two different scanners of the same model were tested, and one scanner acquisition was tested twice with two generations of the same workstation giving six test results in total. The scanners were in active clinical use and are subject to regular maintenance checks. There were three displacement and four velocity results that stood out from the rest and could be regarded as accurate and interchangeable. Among the deformation measurements, three acceptable strain results were found while there were no acceptable strain rate results. Furthermore, the study showed that measurements from scanners of the same model, same acquisition post-processed on different workstations and repeated measurements from the same scanner, can yield disparate results. CONCLUSION: Measurements that are accurate and of interchangeable use can be found for displacement and velocity measurements, but are less likely to be found for strain and strain rate measurements. It is strongly recommended that the ability of each individual US scanner to measure displacement, velocity, strain, and strain rate is evaluated before it is introduced into clinical practice, and it must always be evaluated together with the workstation the scanner is intended to be used in conjunction with.

Ort, förlag, år, upplaga, sidor
OXFORD UNIV PRESS , 2011. Vol. 12, nr 6, s. 467-476
Nyckelord [en]
Ultrasound scanner; Tissue Doppler, Velocity imaging, Strain imaging, Phantom, Quality assurance
Nationell ämneskategori
Medicinsk laboratorieteknik Medicinsk instrumentering
Identifikatorer
URN: urn:nbn:se:kth:diva-34176DOI: 10.1093/ejechocard/jer056ISI: 000293527600012PubMedID: 21565867Scopus ID: 2-s2.0-80051995818OAI: oai:DiVA.org:kth-34176DiVA, id: diva2:419678
Anmärkning

QC 20110527

Tillgänglig från: 2011-05-27 Skapad: 2011-05-27 Senast uppdaterad: 2025-02-10Bibliografiskt granskad
Ingår i avhandling
1. Evaluation of Errors and Limitations in Ultrasound Imaging Systems
Öppna denna publikation i ny flik eller fönster >>Evaluation of Errors and Limitations in Ultrasound Imaging Systems
2011 (Engelska)Doktorsavhandling, sammanläggning (Övrigt vetenskapligt)
Abstract [en]

There are binding regulations requiring safety and efficacy aspects of medical devices. The requirements ask for documentation that the devices are safe and effective for their intended use, i.e. if a device has a measuring function it must be correct. In addition to this there are demands for quality systems describing development, manufacturing, labelling, and manufacturing of a device. The requirements are established to guarantee that non-defective medical devices are used in the routine clinical practice. The fast rates in which the imaging modalities have evolved during the last decades have resulted in numerous new diagnostic tools, such as velocity and deformation imaging in ultrasound imaging. However, it seems as if the development of evaluation methods and test routines has not been able to keep up the same pace. Two of the studies in this thesis, Study I and IV, showed that computed tomography-based and ultrasound based volume measurements can yield very disparate measurements, and that tissue Doppler imaging-based ultrasound measurements can be unreliable.

Furthermore, the new ultrasound modalities impose higher demands on the ultrasound transducers. Transducers are known to be fragile, but defective transducers were less of a problem earlier when the ultrasound systems to a lesser extent were used for measurements. The two other studies, Study II and III, showed that serious transducer errors are very common, and that annual testing of the transducers is not sufficient to guarantee an error free function.

The studies in the thesis indicate that the system with Notified Bodies, in accordance with the EU’s Medical Device Directive, checking the function and manufacturing of medical devices does not work entirely satisfactory. They also show that the evaluation of new methods have led to the undesirable situation, where new measuring tools, such as volume rendering from imaging systems, and tissue Doppler-based velocity and deformation imaging in echocardiography are available for clinicians without proven knowledge about their accuracy.

Ort, förlag, år, upplaga, sidor
Stockholm: KTH Royal Institute of Technology, 2011. s. x, 74
Serie
Trita-STH : report, ISSN 1653-3836 ; 2011:3
Nationell ämneskategori
Medicinsk laboratorieteknik
Identifikatorer
urn:nbn:se:kth:diva-34177 (URN)978-91-7501-026-7 (ISBN)
Disputation
2011-06-10, Sal 3-221, Alfred Nobels Allé 10, Flemingsberg, 13:00 (Svenska)
Opponent
Handledare
Anmärkning
QC 20110527Tillgänglig från: 2011-05-27 Skapad: 2011-05-27 Senast uppdaterad: 2025-02-09Bibliografiskt granskad

Open Access i DiVA

Fulltext saknas i DiVA

Övriga länkar

Förlagets fulltextPubMedScopus

Person

Mårtensson, MattiasBjällmark, AnnaBrodin, Lars-Åke

Sök vidare i DiVA

Av författaren/redaktören
Mårtensson, MattiasBjällmark, AnnaBrodin, Lars-Åke
Av organisationen
Medicinsk bildteknikMedicinsk teknik
I samma tidskrift
European Journal of Echocardiography
Medicinsk laboratorieteknikMedicinsk instrumentering

Sök vidare utanför DiVA

GoogleGoogle Scholar

doi
pubmed
urn-nbn

Altmetricpoäng

doi
pubmed
urn-nbn
Totalt: 419 träffar
RefereraExporteraLänk till posten
Permanent länk

Direktlänk
Referera
Referensformat
  • apa
  • ieee
  • modern-language-association-8th-edition
  • vancouver
  • Annat format
Fler format
Språk
  • de-DE
  • en-GB
  • en-US
  • fi-FI
  • nn-NO
  • nn-NB
  • sv-SE
  • Annat språk
Fler språk
Utmatningsformat
  • html
  • text
  • asciidoc
  • rtf