In vitro contrast-enhanced ultrasound measurements of capillary microcirculation: Comparison between polymer- and phospholipid-shelled microbubbles
KTH, School of Engineering Sciences (SCI), Aeronautical and Vehicle Engineering, Marcus Wallenberg Laboratory MWL.ORCID iD: 0000-0002-3699-396X
KTH, School of Engineering Sciences (SCI), Aeronautical and Vehicle Engineering, MWL Structural and vibroacoustics.ORCID iD: 0000-0001-5760-3919
KTH, School of Technology and Health (STH), Medical Engineering.
CLINTEC, Department of Radiology, Karolinska Institutet.
Show others and affiliations
2011 (English)In: Ultrasonics, ISSN 0041-624X, E-ISSN 1874-9968, Vol. 51, no 1, 40-48 p.Article in journal (Refereed) Published
Abstract [en]

The focus of contrast-enhanced ultrasound research has developed beyond visualizing the blood pool and its flow to new areas such as perfusion imaging, drug and gene therapy, and targeted imaging. In this work comparison between the application of polymer- and phospholipid-shelled ultrasound contrast agents (UCAs) for characterization of the capillary microcirculation is reported. All experiments are carried out using a microtube as a vessel phantom. The first set of experiments evaluates the optimal concentration level where backscattered signal from microbubbles depends on concentration linearly. For the polymer-shelled UCAs the optimal concentration level is reached at a value of about 2 x 10(4) MB/ml, whereas for the phospholipid-shelled UCAs the optimal level is found at about 1 x 10(5) MB/ml.

Despite the fact that the polymer shell occupies 30% of the radius of microbubble, compared to 0.2% of the phospholipid-shelled bubble, approximately 5-fold lower concentration of the polymer UCA is needed for investigation compared to phospholipid-shelled analogues. In the second set of experiments, destruction/replenishment method with varied time intervals ranging from 2 ms to 3 s between destructive and monitoring pulses is employed. The dependence of the peak-to-peak amplitude of backscattered wave versus pulse interval is fitted with an exponential function of the time gamma = A( 1 - exp(-beta t)) where A represents capillary volume and the time constant beta represents velocity of the flow. Taking into account that backscattered signal is linearly proportional to the microbubble concentration, for both types of the UCAs it is observed that capillary volume is linearly proportional to the concentration of the microbubbles, but the estimation of the flow velocity is not affected by the change of the concentration. Using the single capillary model, for the phospholipid-shelled UCA a delay of about 0.2-0.3 s in evaluation of the perfusion characteristics is found while polymer-shelled UCA provide response immediately. The latter at the concentration lower than 3.6 x 10(5) MB/ml have no statistically significant delay (p < 0.01), do not cause any attenuation of the backscattered signal or saturation of the receiving part of the system. In conclusion, these results suggest that the novel polymer-shelled microbubbles have a potential to be used for perfusion evaluation.

Place, publisher, year, edition, pages
2011. Vol. 51, no 1, 40-48 p.
Keyword [en]
Contrast agents, Polymer, Phospholipid, Capillary Perfusion
National Category
Medical Laboratory and Measurements Technologies
URN: urn:nbn:se:kth:diva-12116DOI: 10.1016/j.ultras.2010.05.006ISI: 000282897500006ScopusID: 2-s2.0-77957931904OAI: diva2:302289

manuscript number ULTRAS-D-10-00015 QC20100723 Changed to published

Available from: 2010-03-05 Created: 2010-03-05 Last updated: 2012-08-30Bibliographically approved
In thesis
1. Polymer-shelled Ultrasound Contrast Agents: Characterization and Application
Open this publication in new window or tab >>Polymer-shelled Ultrasound Contrast Agents: Characterization and Application
2010 (English)Doctoral thesis, comprehensive summary (Other academic)
Abstract [en]

Ultrasound-based imaging technique is probably the most used approach for rapid investigationand monitoring of anatomical and physiological conditions of internal organs and tissues.Ultrasound-based techniques do not require the use of ionizing radiation making the tests anexceptionally safe and painless. Operating in the frequency range between 1 to 15 MHz, medicalultrasound provides reliable visual and quantitative information from both superficial structuressuch as muscles and tendons, and also deeper organs such as liver and kidney. From the technicalpoint of view medical ultrasound has a good spatial and temporal resolution. Ultrasound machineis mobile or even portable, which makes it truly bedside modality. And last but not the least,ultrasound investigations are cheaper in comparison to other real time imaging techniques.

Ultrasound imaging techniques can be greatly improved by the use of contrast agents to enhancethe signal from the area of interest (e.g. cardiac or liver tissues) relative to the background.Typically ultrasound contrast agent (UCA) is a suspension of the microbubbles consisting of agas core encapsulated within the solid shell. Generally these devices are injected systemically andfunction to passively enhance the ultrasound echo. In recent years, the UCAs have evolved frombeing just a visualization tool to become a new multifunctional and complex device for drug orgene therapy and targeted imaging.

The overall objective of the project is to test novel polymer shelled microbubbles (MBs) as apossible new generation of ultrasound contrast agents.

During the first year of the project an innovative criterion based on cross-correlation analysis toassess the pressure threshold at which ultrasonic waves fracture the polymer shell of microbubblehas been developed. In addition, acoustic properties of these microbubbles which are relevant totheir use both as contrast agents and drug carriers for localized delivery have been preliminarytested. Furthermore, in order to reconstruct viscoelastic properties of the shell the originalChurch’s model (1995) has been implemented. In collaboration with Karolinska Institutet, imagesof the microbubbles have been acquired with conventional imaging system. These imagesdemonstrate the potential of the novel polymer-shelled microbubbles to be used as contractenhancing agents.

The objective of the second year was to describe the acoustic and mechanical properties ofdifferent types of microbubbles synthesised under varied conditions. This task was divided in twointerrelated parts. In the first part acoustic characterization has been completed in low intensityregion with the study of backscattered power, attenuation and phase velocity. In order torecalculate mechanical properties of the shell existing theoretical model has been furtheriimodified to accommodate the frequency dependence of viscoelastic properties andsimultaneously fit the attenuation and phase velocity data. The results concerning acoustic andmechanical properties of the microbubbles have been sent as a feedback to the manufacture inorder to optimize fabrication protocol for effective image acquisition. In the second part acousticcharacterization has been performed in high intensity region under varied parameters ofexperimental set-up. The results that illustrate the dependence of the fracture pressure thresholdon the system parameters allows us to discuss the potential role of polymer-shelled UCAs as drugcarriers and formulate the protocol for save, localized, cavitation-mediated drug delivery.

For the third year the major task was to move on from the bulk volume in vitro tests towards themicrocapillary study and even further to incorporate the microcapillary into the tissue mimickingultrasound phantom. The last study has the objective to take into account the wave propagationthrough tissue. And last but not the least, the application of the polymer-shelled microbubblesfor evaluation of perfusion characteristics, i.e. capillary volume and velocity of the flow, has beenperformed. Similar tests are carried out with commercially available phospholipid-shelled UCA.Using destruction/replenishment technique it is suggested that the novel polymer-shelledmicrobubbles have a potential for a more accurate perfusion evaluation compared to that ofcommercially available phospholipid-shelled UCA.

In conclusion, proposed polymer-shelled gas-core microbubbles provide a viable system to beused among the next generation of ultrasound contrast agents, which facilitate not only imageenhancement relevant to diagnostics but also localized and specific drug delivery for non-invasivetherapy even in acute conditions.

Place, publisher, year, edition, pages
Stockholm: KTH, 2010. xii, 40 p.
Trita-AVE, ISSN 1651-7660
ultrasound, contrast agents, microbubbles, polymer, perfusion
National Category
Fluid Mechanics and Acoustics Other Engineering and Technologies not elsewhere specified Cardiac and Cardiovascular Systems Cardiac and Cardiovascular Systems
urn:nbn:se:kth:diva-12120 (URN)978-91-7415-572-3 (ISBN)
Public defence
2010-03-26, F3, Lindstedsvägen 26, KTH, 13:00 (English)
“Systems for insitu theranostics using micro-particles triggered by ultra-sound (SIGHT) (FP6-IST-2005-2.5.2, 2006- 2010)
QC20100723Available from: 2010-03-08 Created: 2010-03-05 Last updated: 2010-07-23Bibliographically approved

Open Access in DiVA

No full text

Other links

Publisher's full textScopus

Search in DiVA

By author/editor
Grishenkov, DmitryKari, LeifBrodin, Lars-Åke
By organisation
Marcus Wallenberg Laboratory MWLMWL Structural and vibroacousticsMedical Engineering
In the same journal
Medical Laboratory and Measurements Technologies

Search outside of DiVA

GoogleGoogle Scholar
The number of downloads is the sum of all downloads of full texts. It may include eg previous versions that are now no longer available

Altmetric score

Total: 87 hits
ReferencesLink to record
Permanent link

Direct link