2009 (English)Article in journal (Refereed) Published
Abstract [en]

Electrodes coated with the conducting polymer poly(3,4-ethylene dioxythiophene) (PEDOT) possess attractive electrochemical properties for stimulation or recording in the nervous system. Biomolecules, added as counter ions in electropolymerization, could further improve the biomaterial properties, eliminating the need for surfactant counter ions in the process. Such PEDOT/biomolecular composites, using heparin or hyaluronic acid, have previously been investigated electrochemically. In the present study, their biocompatibility is evaluated. An agarose overlay assay using L929 fibroblasts, and elution and direct contact tests on human neuroblastoma SH-SY5Y cells are applied to investigate cytotoxicity in vitro. PEDOT: heparin was further evaluated in vivo through polymer-coated implants in rodent cortex. No cytotoxic response was seen to any of the PEDOT materials tested. The examination of cortical tissue exposed to polymer-coated implants showed extensive glial scarring irrespective of implant material (Pt:polymer or Pt). However, quantification of immunological response, through distance measurements from implant site to closest neuron and counting of ED1+ cell density around implant, was comparable to those of platinum controls. These results indicate that PEDOT: heparin surfaces were non-cytotoxic and show no marked difference in immunological response in cortical tissue compared to pure platinum controls.

Place, publisher, year, edition, pages
2009. Vol. 4, no 4
Keyword [en]
silicon microelectrode arrays, conducting polymer, hyaluronic-acid, in-vitro, surface modification, fibrin gels, poly(3, 4-ethylenedioxythiophene), biocompatibility, probes, regeneration
URN: urn:nbn:se:kth:diva-18682DOI: 10.1088/1748-6041/4/4/045009ISI: 000268871000018ScopusID: 2-s2.0-69049098145OAI: diva2:336729

QC 20100525

Available from: 2010-08-05 Created: 2010-08-05 Last updated: 2014-09-24Bibliographically approved

Open Access in DiVA

No full text

Other links

Publisher's full textScopus

Search in DiVA

By author/editor
Asplund, MariaThaning, Elinvon Holst, Hans
By organisation
Neuronic Engineering (Closed 20130701)

Search outside of DiVA

GoogleGoogle Scholar
The number of downloads is the sum of all downloads of full texts. It may include eg previous versions that are now no longer available

Altmetric score

Total: 54 hits
ReferencesLink to record
Permanent link

Direct link