An evaluation and improvement of an in vitro heart phantom of the hearts coronary circulation.
KTH, School of Technology and Health (STH).
2013 (English)Independent thesis Advanced level (degree of Master (Two Years)), 20 credits / 30 HE creditsStudent thesis
Abstract [en]

The purpose of this thesis has been to validate the in-vitro heart simulation model of the coronary arteries called Flowlab, identify limitations and potential problems, and to offer suggestions for improvement. Flowlab emulates arterial characteristics such as pressure and flow, providing a simulation environment with the ability to measure the volumetric coronary blood flow and arterial blood pressure. Compared to current simulation systems, this will give a better understanding of both position and severity of the cardiovascular disease, and also provide essential information regarding the hemodynamics in the coronary arteries.

 

To understand the fluid dynamics of the coronary system and gain a deeper understanding of the arterial function and physiology, the first phase of this thesis will focus on gathering information regarding the anatomy, physiology and hemodynamics of the coronary arteries. This will also be done to provide the Flowlab construction with appropriate measurements for the coronary simulation pipe, and input data for its final design.

 

The Flowlab construction will then be tested to verify the accuracy of the produced values compared with the sought after values of pressure and flow, to further enable an adjustment of the system.

 

The results from the test show that the final calibration of the simulation environment was not sufficient to produce an accurate result for both pressure and flow regulation, only flow. The flow had an accuracy of 10 %, while the generated pressure was underestimated at low pressures and overestimated at high pressure at up to 20 %. An accuracy of 10 % for generated flow is satisfying and sufficient; meanwhile the pressure calibration had to be altered for more reliable data.

 

The analysis also revealed several shortages in the design of the Flowlab system that needed to be adjusted for more consistent pressure and flow measurements.

Place, publisher, year, edition, pages
2013. , 85 p.
Series
TRITA-STH, 2013:118
National Category
Medical Engineering
Identifiers
URN: urn:nbn:se:kth:diva-133775OAI: oai:DiVA.org:kth-133775DiVA: diva2:663278
External cooperation
St Jude Medical
Subject / course
Medical Engineering
Educational program
Master of Science in Engineering - Medical Engineering
Presentation
2013-06-12, 10:54 (English)
Supervisors
Examiners
Available from: 2014-03-05 Created: 2013-11-11 Last updated: 2014-03-05Bibliographically approved

Open Access in DiVA

SAFIA_BENNANI_STUDENT_THESES(1548 kB)182 downloads
File information
File name FULLTEXT01.pdfFile size 1548 kBChecksum SHA-512
e7797730e0c71c2a8cdfc61171c096e6dc71c9d42df60357f7d8a83efdacca3d23d3ef2559168c160c845f0b0525fff5c4e3d2d2d3d66c19d37f29b852eaaeb4
Type fulltextMimetype application/pdf

By organisation
School of Technology and Health (STH)
Medical Engineering

Search outside of DiVA

GoogleGoogle Scholar
Total: 182 downloads
The number of downloads is the sum of all downloads of full texts. It may include eg previous versions that are now no longer available

urn-nbn

Altmetric score

urn-nbn
Total: 316 hits
CiteExportLink to record
Permanent link

Direct link
Cite
Citation style
  • apa
  • harvard1
  • ieee
  • modern-language-association-8th-edition
  • vancouver
  • Other style
More styles
Language
  • de-DE
  • en-GB
  • en-US
  • fi-FI
  • nn-NO
  • nn-NB
  • sv-SE
  • Other locale
More languages
Output format
  • html
  • text
  • asciidoc
  • rtf