Ultrasound Contrast Agents: Fabrication, size distribution and visualization
KTH, School of Technology and Health (STH), Medical Engineering, Medical Imaging. (Contrast Enhanced Medical Imaging)
2011 (English)Independent thesis Advanced level (degree of Master (Two Years)), 20 credits / 30 HE creditsStudent thesis
Abstract [en]

Ultrasound contrast agents composed of micro-bubble filled with gas are introduced to increase the backscattered power from blood. Their intravenously injection results in the improved contrast in the images.

The aim of this master thesis project is to manufacture MB suspension at varied temperature and shear forces and to inspect the size distribution and concentration of the PVA-shelled micro-bubble with standard methods according to the developed protocol. A pulser-receiver (Panametrics PR 5072) setup combined with two transducers (2.25 MHz and 5 MHz) was used to investigate the backscattered enhancement of the micro-bubble suspension.

Images were collected with transmission optical microscope (OLYMPUS IX71) with the aid of counting chamber. The diameter and concentration of the micro-bubbles were analyzed by Image J. The pulser-receiver setup was used to test the acoustic response.

The mean diameter of micro-bubbles was from 2.03 to 4.38 µm with a standard deviation between 0.40 and 1.12 µm and the micro-bubble concentration varied from 0.07× to 5.22× MBs/ml. The enhancement of the ultrasound backscattered power was greater than 20 dB or even reached 30 dB when the energy was increased. 

Place, publisher, year, edition, pages
2011. , 41 p.
Keyword [en]
Ultrasound, contrast agent, microbubbles
National Category
Medical Materials Medical Equipment Engineering
URN: urn:nbn:se:kth:diva-152586OAI: diva2:750462
Subject / course
Medical Engineering
Educational program
Master of Science - Medical Engineering
Available from: 2014-10-13 Created: 2014-09-29 Last updated: 2014-10-13Bibliographically approved

Open Access in DiVA

Ultrasound Contrast Agents: Fabrication, size distribution and visualization(1283 kB)112 downloads
File information
File name FULLTEXT01.pdfFile size 1283 kBChecksum SHA-512
Type fulltextMimetype application/pdf

By organisation
Medical Imaging
Medical MaterialsMedical Equipment Engineering

Search outside of DiVA

GoogleGoogle Scholar
Total: 112 downloads
The number of downloads is the sum of all downloads of full texts. It may include eg previous versions that are now no longer available

Total: 107 hits
ReferencesLink to record
Permanent link

Direct link