The influence of anatomical noise on optimal beam quality in mammography
KTH, School of Engineering Sciences (SCI), Physics.
KTH, School of Engineering Sciences (SCI), Physics.
2014 (English)In: Medical physics (Lancaster), ISSN 0094-2405, Vol. 41, no 12, 121903- p.Article in journal (Refereed) Published
Abstract [en]

Purpose: Beam-quality optimization in digital mammography traditionally considers detection of a target obscured by quantum noise in a homogeneous background. This does not correspond well to the clinical imaging task because real mammographic images contain a complex superposition of anatomical structures, resulting in anatomical noise that may dominate over quantum noise. The purpose of this paper is to assess the influence on optimal beam quality in mammography when anatomical noise is taken into account. Methods: The detectability of microcalcifications and masses was quantified using a theoretical ideal-observer model that included quantum noise as well as anatomical noise and a simplified model of a photon-counting mammography system. The outcome was experimentally verified using two types of simulated tissue phantoms. Results: The theoretical model showed that the detectability of tumors and microcalcifications behaves differently with respect to beam quality and dose. The results for small microcalcifications were similar to what traditional optimization methods yield, which is to be expected because quantum noise dominates over anatomical noise at high spatial frequencies. For larger tumors, however, low-frequency anatomical noise was the limiting factor. Because anatomical structure noise has similar energy dependence as tumor contrast, the optimal x-ray energy was found to be higher and the useful energy region was wider than traditional methods suggest. A simplified scalar model was able to capture this behavior using a fitted noise mixing parameter. The phantom measurements confirmed these theoretical results. Conclusions: It was shown that since quantum noise constitutes only a small fraction of the noise, the dose could be reduced substantially without sacrificing tumor detectability. Furthermore, when anatomical noise is included, the tube voltage can be increased well beyond what is conventionally considered optimal and used clinically, without loss of image quality. However, no such conclusions can be drawn for the more complex mammographic imaging task as a whole. (C) 2014 American Association of Physicists in Medicine.

Place, publisher, year, edition, pages
2014. Vol. 41, no 12, 121903- p.
Keyword [en]
mammography, optimization, anatomical noise, ideal observer
National Category
Radiology, Nuclear Medicine and Medical Imaging
URN: urn:nbn:se:kth:diva-159026DOI: 10.1118/1.4900611ISI: 000346176300019ScopusID: 2-s2.0-84910656674OAI: diva2:785709

QC 20150203

Available from: 2015-02-03 Created: 2015-01-20 Last updated: 2015-02-03Bibliographically approved

Open Access in DiVA

No full text

Other links

Publisher's full textScopus

Search in DiVA

By author/editor
Cederström, BjörnFredenberg, Erik
By organisation
In the same journal
Medical physics (Lancaster)
Radiology, Nuclear Medicine and Medical Imaging

Search outside of DiVA

GoogleGoogle Scholar
The number of downloads is the sum of all downloads of full texts. It may include eg previous versions that are now no longer available

Altmetric score

Total: 12 hits
ReferencesLink to record
Permanent link

Direct link