Central and peripheral refraction measured by a novel double-pass instrumentVise andre og tillknytning
2023 (engelsk)Inngår i: Biomedical Optics Express, ISSN 2156-7085, Vol. 14, nr 6Artikkel i tidsskrift (Fagfellevurdert) Published
Abstract [en]
A novel double-pass instrument and its data analysis method for the measurement of central and peripheral refraction is presented and validated in a group of healthy subjects. The instrument acquires in-vivo, non-cycloplegic, double-pass, through-focus images of the eye’s central and peripheral point-spread function (PSF) using an infrared laser source, a tunable lens and a CMOS camera. The through-focus images were analyzed to determine defocus and astigmatism at 0° and 30° visual field. These values were compared to those obtained with a lab-based Hartmann-Shack wavefront sensor. The two instruments provided data showing good correlation at both eccentricities, particularly in the estimation of defocus.
sted, utgiver, år, opplag, sider
Optica Publishing Group , 2023. Vol. 14, nr 6
HSV kategori
Identifikatorer
URN: urn:nbn:se:kth:diva-328444DOI: 10.1364/boe.489881ISI: 001014836500007PubMedID: 37342694Scopus ID: 2-s2.0-85162035591OAI: oai:DiVA.org:kth-328444DiVA, id: diva2:1764823