New opportunities for time-resolved imaging using diffraction-limited storage ringsVisa övriga samt affilieringar
2024 (Engelska)Ingår i: Journal of Synchrotron Radiation, ISSN 0909-0495, E-ISSN 1600-5775, Vol. 31, nr Pt 5, s. 1299-1307Artikel i tidskrift (Refereegranskat) Published
Abstract [en]
The advent of diffraction-limited storage rings (DLSRs) has boosted the brilliance or coherent flux by one to two orders of magnitude with respect to the previous generation. One consequence of this brilliance enhancement is an increase in the flux density or number of photons per unit of area and time, which opens new possibilities for the spatiotemporal resolution of X-ray imaging techniques. This paper studies the time-resolved microscopy capabilities of such facilities by benchmarking the ForMAX beamline at the MAX IV storage ring. It is demonstrated that this enhanced flux density using a single harmonic of the source allows micrometre-resolution time-resolved imaging at 2000 tomograms per second and 1.1 MHz 2D acquisition rates using the full dynamic range of the detector system.
Ort, förlag, år, upplaga, sidor
International Union of Crystallography , 2024. Vol. 31, nr Pt 5, s. 1299-1307
Nyckelord [en]
diffraction-limited storage rings, ForMAX beamline, MAX IV, megahertz imaging, time-resolved microscopy
Nationell ämneskategori
Subatomär fysik Atom- och molekylfysik och optik
Identifikatorer
URN: urn:nbn:se:kth:diva-353445DOI: 10.1107/S1600577524005290ISI: 001362324700035Scopus ID: 2-s2.0-85203165153OAI: oai:DiVA.org:kth-353445DiVA, id: diva2:1899120
Anmärkning
QC 20240924
2024-09-192024-09-192025-12-05Bibliografiskt granskad