A Mixer-First Analog Front-End for Dry-Electrode Bioimpedance Spectroscopy
KTH, School of Electrical Engineering and Computer Science (EECS), Electrical Engineering, Electronics and Embedded systems.ORCID iD: 0000-0001-7549-0858
KTH, School of Electrical Engineering and Computer Science (EECS), Electrical Engineering, Electronics and Embedded systems.ORCID iD: 0000-0003-3802-7834
2023 (English)In: BioCAS 2023 - 2023 IEEE Biomedical Circuits and Systems Conference, Conference Proceedings, Institute of Electrical and Electronics Engineers (IEEE) , 2023Conference paper, Published paper (Refereed)
Abstract [en]

This paper presents a high input impedance, low-noise, and low-distortion analog front-end (AFE) for bioimpedance (bio-Z) spectroscopy measurements targeting neuromuscular health assessments. The proposed 8-phase quadrature mixer-first architecture achieves a high input impedance through passive mixers driven by non-overlapping clocks. The 8-phase signals are recombined to extract the real and imaginary parts of the bio-Z, while rejecting unwanted harmonics to improve linearity. Programmability of the AFE enables accurate bio-Z measurements up to 10 kΩ for 11 logarithmically spaced frequencies, in the 1 kHz to 1 MHz frequency range. The AFE, designed in a 0.18 μm CMOS process, consumes 245.99 μW at the lowest gain and 300.56 μW at the highest gain, and occupies 2.4 mm2 silicon area. Post-layout simulation results show that the input impedance is always higher than the electrode impedance by more than 10x. The AFE achieves a sensitivity of 7.7 mΩrms, and a maximum SNDR of 103.87 dBFS over a 61 Hz bandwidth. These results demonstrate that the proposed AFE enables bio-Z measurements, using dry electrodes, with errors below 1%.

Place, publisher, year, edition, pages
Institute of Electrical and Electronics Engineers (IEEE) , 2023.
Keywords [en]
analog front-end, bioimpedance spectroscopy, dry electrodes, high input impedance, low-distortion, low-noise, passive mixer, programmable gain
National Category
Other Electrical Engineering, Electronic Engineering, Information Engineering
Research subject
Electrical Engineering; Medical Technology
Identifiers
URN: urn:nbn:se:kth:diva-343702DOI: 10.1109/BioCAS58349.2023.10388966Scopus ID: 2-s2.0-85184898068OAI: oai:DiVA.org:kth-343702DiVA, id: diva2:1839897
Conference
2023 IEEE Biomedical Circuits and Systems Conference, BioCAS 2023, Toronto, Canada, Oct 19 2023 - Oct 21 2023
Funder
Swedish Foundation for Strategic Research, ITM17-0079
Note

QC 20240223

Part of ISBN 979-8-3503-0026-0

Available from: 2024-02-22 Created: 2024-02-22 Last updated: 2024-09-02Bibliographically approved
In thesis
1. Fully Integrated Bioimpedance Spectroscopy Interface for Wearable Electrical Impedance Myography
Open this publication in new window or tab >>Fully Integrated Bioimpedance Spectroscopy Interface for Wearable Electrical Impedance Myography
2024 (English)Doctoral thesis, comprehensive summary (Other academic)
Abstract [en]

Multi-Frequency surface Electrical Impedance Myography (MF-sEIM) is a technique that provides valuable electrophysiological information of muscles. This technique measures bio-Z spectroscopy of muscles, and applies Ohm's law by injecting a small-amplitude and high-frequency current into tissues and measuring the voltage response. As biological tissues are electrolytic conductors, this technique provides information on fundamental dielectric properties of tissues, which are an objective biomarker of neuromuscular disorders and have practical value in several muscle healthcare applications. Due to its versatility, simplicity, and ease of integration, this technique is a great candidate to complement stand-alone surface electromyography (sEMG) in wearable devices for continuous monitoring of muscle health. Nonetheless, wearable MF-sEIM imposes challenging requirements on the bio-Z spectroscopy interface. Previously reported solutions fall short of meeting these requirements, or do so with low power efficiency.

This thesis focuses on the research and development of a fully integrated bio-Z spectroscopy interface, which complies with the challenging requirements of wearable MF-sEIM in a power-efficient way. The electrophysiological mechanisms of MF-sEIM and the system-level requirements for clinical relevance were investigated, as MF-sEIM is a relatively novel technique, which requires standarization. From this established set of requirements, the main building blocks of the \mbox{bio-Z} spectroscopy interface, i.e., the current signal generator (CSG), and voltage readout, were developed. The CSG generates pseudo-sine waves through direct digital synthesis (DDS) to obtain the required linearity with high power efficiency. A high-linearity full current-mode CSG was also proposed to comply with the stricter bio-Z accuracy requirements of clinical diagnostics. The voltage readout is based on a low-IF quadrature (I/Q) demodulation architecture and features a pseudo 2-path bandpass (BP) Delta-Sigma ADC to achieve high precision and power-to-noise efficiency. A mixer-first analog front-end (AFE) was also proposed to enable bio-Z spectroscopy measurements employing dry electrodes. The bio-Z interface was integrated with a 16-Channel sEMG AFE and a 4-Channel neuromuscular stimulator (NMES) in an application-specific integrated circuit (ASIC). Experimental results show that the implemented bio-Z spectroscopy interface achieves a comparable performance with the state of the art, while being capable of detecting the large baseline and the time-varying impedances of muscle. A proof-of-concept system, based on the multi-modal ASIC, was developed. This system demonstrates the potential of combining real-time monitoring of MF-sEIM and sEMG for detecting muscle fatigue, enabling efficient closed-loop NMES.

Abstract [sv]

Multifrekvent ytelektroimpedansmyografi (MF-sEIM) är en teknik som ger värdefull elektrofysiologisk information om muskler. Denna teknik mäter bioimpedans-spektroskopi (bio-Z-spektroskpi) av muskler och tillämpar Ohms lag genom att injicera en högfrekvent småsignalström i vävnader och mäta spänningssvaret. Eftersom biologiska vävnader är elektrolytiska ledare, ger denna teknik information om grundläggande dielektriska egenskaper hos vävnader, som är en objektiv biomarkör för neuromuskulära störningar och har praktiskt värde i flera tillämpningar för muskelsjukvård. På grund av dess mångsidighet och enkel integration är denna teknik en utmärkt kandidat för att komplettera fristående ytelektromyografi (sEMG) i bärbara enheter för kontinuerlig övervakning av muskelhälsa. Icke desto mindre ställer bärbar MF-sEIM utmanande krav på bio-Z-spektroskopigränssnittet. Tidigare rapporterade lösningar uppfyller inte dessa krav, eller gör det med låg energieffektivitet.

Denna avhandling fokuserar på forskning och utveckling av ett helt integrerat bio-Z-spektroskopigränssnitt, som uppfyller de utmanande kraven för bärbar MF-sEIM på ett energieffektivt sätt. De elektrofysiologiska mekanismerna för MF-sEIM och kraven på systemnivå för klinisk relevans undersöktes, eftersom MF-sEIM är en relativt ny teknik som kräver standardisering. Från de definerade kraven på systemnivå utvecklades huvudbyggstenarna för bio-Z-spektroskopigränssnittet, dvs. strömsignalgeneratorn (CSG) och spänningsavläsningen. CSG genererar pseudo-sinusvågor genom direkt digital syntes (DDS) för att erhålla den erforderliga linjäriteten med hög energieffektivitet. En CSG i fullströmsläge med hög linjäritet föreslogs också för att uppfylla de striktare noggrannhetskraven på bioimpedansmätning för klinisk diagnostik. Spänningsavläsningen är baserad på en demoduleringsarkitektur med låg IF-kvadratur (I/Q) och har en pseudo 2-vägs bandpass (BP) Delta-Sigma ADC för att uppnå hög precision och energi till brus effektivitet. En mixer-framför analog front-end (AFE) föreslogs också för att möjliggöra bio-Z-spektroskopimätningar med användning av torra elektroder. Bio-Z-gränssnittet integrerades med en 16-kanals sEMG AFE och en 4-kanals neuromuskulär stimulator (NMES) i en applikationsspecifik integrerad krets (ASIC). Experimentella resultat visar att det implementerade bio-Z-spektroskopigränssnittet uppnår en prestanda som är jämförbar med den senaste tekniken, samtidigt som den kan detektera storsignalsimpedansen hos muskler och dess tidsvariation. Ett prototyp, baserat på den multimodala ASIC, utvecklades. Detta system visar potentialen i att kombinera realtidsövervakning av MF-sEIM och sEMG för att upptäcka muskeltrötthet, vilket möjliggör effektiva NMES med återkoppling.

Place, publisher, year, edition, pages
Stockholm: KTH Royal Institute of Technology, 2024. p. xix, 84
Series
TRITA-EECS-AVL ; 2024:62
Keywords
Musculoskeletal health, bioimpedance (bio-Z) spectroscopy, electrical impedance myography (EIM), biomedical modeling, ASIC, ultra-low power, high-precision, sinusoidal signal generator, readout, analog front-end, bandpass delta-sigma ADC, multi-modal sensing., Muskuloskeletal hälsa, bioimpedans-spectroscopi (bio-Z-spektroskpi), elektroimpedansmyografi (EIM), biomedicinsk modellering, ASIC, ultralåg energi, hög precision, sinusformad signalgenerator, avläsning, analog front-end, bandpass delta-sigma ADC, multimodal avkänning.
National Category
Electrical Engineering, Electronic Engineering, Information Engineering
Research subject
Information and Communication Technology
Identifiers
urn:nbn:se:kth:diva-352340 (URN)978-91-8106-022-5 (ISBN)
Public defence
2024-09-27, https://kth-se.zoom.us/j/2593490856, Ka-Sal C, Electrum, Kistagången 16, Kista, 13:00 (English)
Opponent
Supervisors
Funder
Swedish Foundation for Strategic Research, ITM17-0079
Note

QC 20240829

Available from: 2024-08-29 Created: 2024-08-28 Last updated: 2026-01-13Bibliographically approved

Open Access in DiVA

fulltext(656 kB)72 downloads
File information
File name FULLTEXT02.pdfFile size 656 kBChecksum SHA-512
3ff71e3685dee9ab5d6c21a3229907f3dab8fa6f39937ba49c7ce40cc66a0d350662b5c118a21967caabb70408a3846013ec233cca3a5cec489d0c550e72eeae
Type fulltextMimetype application/pdf

Other links

Publisher's full textScopus

Authority records

Fernández Schrunder, AlejandroRusu, Ana

Search in DiVA

By author/editor
Fernández Schrunder, AlejandroRusu, Ana
By organisation
Electronics and Embedded systems
Other Electrical Engineering, Electronic Engineering, Information Engineering

Search outside of DiVA

GoogleGoogle Scholar
Total: 72 downloads
The number of downloads is the sum of all downloads of full texts. It may include eg previous versions that are now no longer available

doi
urn-nbn

Altmetric score

doi
urn-nbn
Total: 258 hits
CiteExportLink to record
Permanent link

Direct link
Cite
Citation style
  • apa
  • ieee
  • modern-language-association-8th-edition
  • vancouver
  • Other style
More styles
Language
  • de-DE
  • en-GB
  • en-US
  • fi-FI
  • nn-NO
  • nn-NB
  • sv-SE
  • Other locale
More languages
Output format
  • html
  • text
  • asciidoc
  • rtf