Cross-calibration and first vertical ECE measurement of electron energy distribution in the TCV tokamakShow others and affiliations
2024 (English)In: Plasma Physics and Controlled Fusion, ISSN 0741-3335, E-ISSN 1361-6587, Vol. 66, no 12, article id 125010Article in journal (Refereed) Published
Abstract [en]
This paper describes the first vertical electron cyclotron emission measurement of non-thermal electron distributions in the Tokamak & agrave; Configuration Variable. These measurements were conducted in runaway electron scenarios and in the presence of electron cyclotron current drive. Measured intensities of linearly polarized X- and O-mode radiation from fast electrons allow the analysis of the energy distribution. The measurements were made possible through the creation of an operational regime for the diagnostic that is free of thermal background radiation, in relaxed electron density operations. This operational regime notably enables the cross-calibration of the diagnostic system, relying on thermal plasma measurements and modeling with the ray-tracing code SPECE.
Place, publisher, year, edition, pages
IOP Publishing , 2024. Vol. 66, no 12, article id 125010
Keywords [en]
electron cyclotron emission, non-thermal electron distribution, runaway electron, electron cyclotron current drive, radiometry, calibration, synthetic diagnostics
National Category
Fusion, Plasma and Space Physics
Identifiers
URN: urn:nbn:se:kth:diva-356496DOI: 10.1088/1361-6587/ad88a0ISI: 001347954100001Scopus ID: 2-s2.0-85209398036OAI: oai:DiVA.org:kth-356496DiVA, id: diva2:1913680
Note
QC 20241115